Principle and Application

This kit adopts the method of indirect competitive enzyme-linked immunoassay (ELISA) to detect Sulfamethoxazole (SMZ) in the sample such as tissue, serum, honey, milk and urine. The kit is composed of Microtiter Plate coated with coupled antigens, SMZ-HRP enzyme conjugates, antibodies, standards and other supporting reagents. During the detection, with adding standards or samples, the SMZ in the samples will compete with the coupled antigens to combine with anti-SMZ antibodies. After adding enzyme conjugates, take coloration with TMB substrates. Absorbance value of the samples is a negative correlation with SMZ content. Lastly, comparing with the standard curve, the obtained concentration is multiplied by the sample dilution ratio. SMZ residues in sample can be concluded.

Technique Data

>Kit Sensitivity: 0.1ppb (ng/mL)

>Reactive Mode: 25℃, 45min~15min

>Detection Limits:

Sample

Detection Limits

Tissue (higher detection limit)

0.1ppb

Tissue (lower detection limit)

1ppb

Serum, urine

0.4ppb

Honey

0.1ppb

Milk

2ppb

Feed

4ppb

Eggs

0.2ppb

>Cross-reaction Rate:

Drug name

Cross-reaction Rate

Sensitivity

/ppb

Sulfamethoxazole (SMZ)

100%

0.1


>Sample Recovery Rate:

Sample

Recovery Rate

Tissue, honey,eggs

85±25%

Urine, milk, serum, feed

80±25%

Composition of the Kit

Reagent

Specification

Microtiter Plate

8wells× 12strips

Standard: 0ppb, 0.1ppb, 0.3ppb, 0.9ppb, 2.7ppb, 8.1ppb

1×1.0mL

High Standard1ppm(red cap)

1×1.0mL

Antibody solution (blue cap)

1×5.5mL

Enzyme conjugate (red cap)

1×5.5mL

Substrate Reagent A (white cap)

1×6mL

Substrate Reagent B (black cap)

1×6mL

Stop Solution (yellow cap)

1×6mL

Concentrated Wash Buffer (20×)(white cap)

1×40mL

Redissolving Solution(2×) (yellow cap)

1×50mL

Instructions

1

Adhesive Membrane

1

Sealed bag

1