Principle and Application

This kit adopts the method of indirect competitive enzyme-linked immunoassay (ELISA) to detect Clenbuterol (CLE) in the sample such as urine, tissue and feed. The kit is composed of Microtiter Plate coated with coupled antigens, CLE-HRP enzyme conjugates, antibodies, standards and other supporting reagents. During the detection, with adding standards or samples, the CLE in the samples will compete with the coupled antigens to combine with anti-CLE antibodies. After adding enzyme conjugates, take coloration with TMB substrates. Absorbance value of the samples is a negative correlation with CLE content. Lastly, comparing with the standard curve, the obtained concentration is multiplied by the sample dilution ratio. CLE residues in sample can be concluded.

Technique Data

>Kit Sensitivity: 0.05ppb (ng/mL)

>Reactive Mode: 25℃, 30min~15min

>Detection Limits:

Sample

Detection Limits

Urine

0.1ppb

Tissue

0.2ppb

Feed

1ppb

Pork liver

0.5ppb

>Cross-reaction Rate:

Clenbuterol………………………………………………100%

Terbutalin………………………………………………. <1%

Mabuterol………………………………………………. .<1%

Brombuterol……………………………………………...<1%

Salbutamol……………………………………………….<1%

Ractopamine…………………………………………….<1%

>Sample Recovery Rate:

Sample

Recovery Rate

Urine

95±10%

Tissue, feed, Pork liver

85±15%

Composition of the Kit

Reagent

Specification

Microtiter Plate

8wells× 12strips

Standard: 0ppb, 0.05ppb, 0.15ppb, 0.45ppb, 1.35ppb,4.05ppb

1×1.0mL

High Standard100ppb(red cap)

1×1.0mL

Antibody solution (blue cap)

1×5.5mL

Enzyme conjugate (red cap)

1×5.5mL

Substrate Reagent A (white cap)

1×6mL

Substrate Reagent B (black cap)

1×6mL

Stop Solution (yellow cap)

1×6mL

Concentrated Wash Buffer (20×)(white cap)

1×40mL

Concentrated Redissolving Solution(10×)(yellow cap)

1×50mL

Instructions

1

Adhesive Membrane

1

Sealed bag

1